Submit Manuscript  

Article Details


An Active and Low-Cost Microwave Imaging System for Detection of Breast Cancer Using Back Scattered Signal

Author(s):

Anupma Gupta*, Paras Chawla, Ankush Kansal and Kulbir Singh   Pages 1 - 16 ( 16 )

Abstract:


A defected ground antenna with dielectric reflector is designed and investigated for breast tumour diagnosis. Ultra-wide band resonance (3.1 to 10.6 GHz) is achieved by etching two slots and adding a narrow vertical strip in a patch antenna. A high dielectric constant substrate is added below the antenna, which shows remarkable effect on performance. Antenna performance is verified experimentally on an artificially fabricated breast tissue and tumour. Malignant tissue has different dielectric properties than the normal tissue, that causes deviation in the scattered antenna power. Average value of backscattered signal variation and ground penetrating radar (GPR) algorithm is used to localize the tumour of radius 4mm in breast tissue.

Keywords:

Microwave Imaging, ultrawideband, dielectric reflector, GPR algorithm, breast tumour detection, backscattered signal

Affiliation:

Department of Electronics and Communication Engineering, Thapar institute of Engineering and Technology, Patiala, Department of Electronics and Communication Engineering, Chandigarh University, Department of Electronics and Communication Engineering, Thapar institute of Engineering and Technology, Patiala, Department of Electronics and Communication Engineering, Thapar institute of Engineering and Technology, Patiala



Read Full-Text article